Designed for your clinical, diagnostic and research needs.
Product Selection
Model Replacement Plug Type Capacity Electrical RGL1204A TSX1205GA NEMA 5-15P 11.5 cu. ft., 326L 115V 60Hz RGL1204D TSX1205GD NEMA 6-15P 11.5 cu. ft., 326L 208/230V 60Hz RGL2304A TSX2305GA NEMA 5-15P 23.3 cu. ft., 659L 115V 60Hz RGL2304D TSX2305GD NEMA 6-15P 23.3 cu. ft., 659L 208/230V 60Hz RGL3004A TSX3005GA NEMA 5-15P 29.2 cu. ft., 826L 115V 60Hz RGL3004D TSX3005GD NEMA 6-15P 29.2 cu. ft., 826L 208/230V 60Hz REL4504A TSX4505GA NEMA 5-20P 45.8 cu. ft., 1297L 115V 60Hz REL4504D TSX4505GD NEMA 6-15P 45.8 cu. ft., 1297L 208/230V 60Hz RGL5004A TSX5005GA NEMA 5-20P 51.1 cu. ft., 1447L 115V 60Hz RGL5004D TSX5005GD NEMA 6-15P 51.1 cu. ft., 1447L 208/230V 60Hz RGL7504A TSG72RPGA NEMA 5-20P 78.8 cu. ft., 2231L 115V 60Hz RGL7504D NEMA 6-15P 78.8 cu. ft., 2231L 208/230V 60Hz
Description
- Fulfill a variety of space and storage requirements using Thermo Scientific Revco High-Performance RGL Series Laboratory Refrigerators with Glass Doors.
- These refrigerators feature a microprocessor control system, positive airflow systems, industrial-quality cabinet construction and extra-strength refrigeration compressors that are ideal for industrial, clinical and scientific applications.
Features
- Adjustable temperatures between 1° and 8°C (factory preset at 4°C).
- Automatic defrost and condensate removal.
- Microprocessor control.
- Digital electronic technology designed to simplify operation and deliver greater accuracy.
- Icon-based controls.
- Large, easy-to-read digital display of temperature within 0.1°C.
- Graphic thermometer is steady green when temperature is normal, flashes top bar for overtemperature, flashes bulb for undertemperature.
- Sensor automatically initiates defrost cycle on demand, eliminating need for a defrost timer, minimizing frost buildup, and optimizing coil efficiency.
Alarm System
- Audible/visual warnings for temperature deviations or power failure.
- Key-operated ON/OFF/ALARM ON master switch for setpoint security.
- Push-button alarm test and remote alarm contacts.
- Alarm silence, ringback, and auto reset.
- Controller battery backup system.
- Alarm setpoints are preset to warn of temperature deviation at 5.5°C overtemperature and 1.5°C undertemperature.
Construction
- Positive forced-air circulation systems.
- Industrial-grade cabinet construction and compressors.
- Self-closing, key-locked doors.
- Epoxy-coated, open-wire shelves adjustable in 1" (2.5cm) increments.
- Cold-rolled steel interior and exterior finished with high-impact powder paint coating.
- Interior lights with door-activated ON/OFF switch and independent switch on control panel.
- Adjustable leveling feet (except 4.9 cu. ft. undercounter model).
- 45.8 cu. ft. (1297L) units have double sliding glass doors, base-mounted controls and refrigeration systems, and dual-wheel 2" (5cm) casters instead of leveling feet.
- Undercounter model has a 1" (2.5cm) dia. access port with plug.
Ordering Information
- Alarm systems, temperature chart recorders, and power monitors are available separately.
- Other voltages are available; contact our Customer Service Department for more information.
Warranty
- Valid in the USA only.