Designed to meet strict requirements established by the AABB and FDA for storage of whole blood and blood components.

Product Selection

Model Replacement Door Style Capacity Plug Type Electrical
REB404A   Single, Solid 4.9 cu. ft., 139L NEMA 5-15P 115V 60Hz
REB404D   Single, Solid 4.9 cu. ft., 139L NEMA 6-15P 208/230V 60Hz
REB1204A TSX1204BA Single, Glass 11.5 cu. ft., 326L NEMA 5-15P 115V 60Hz
REB1204D TSX1204BD Single, Glass 11.5 cu. ft., 326L NEMA 6-15P 208/230V 60Hz
REB2304A TSX2304BA Single, Glass 23.3 cu. ft., 659L NEMA 5-15P 115V 60Hz
REB2304D TSX2304BD Single, Glass 23.3 cu. ft., 659L NEMA 6-15P 208/230V 60Hz
REB3004A TSX3004BA Single, Glass 29.2 cu. ft., 826L NEMA 5-15P 115V 60Hz
REB3004D TSX3004BD Single, Glass 29.2 cu. ft., 826L NEMA 6-15P 208/230V 60Hz
REB5004A TSX5004BA Double, Glass 51.1 cu. ft., 1447L NEMA 5-20P 115V 60Hz
REB5004D TSX5004BD Double, Glass 51.1 cu. ft., 1447L NEMA 6-15P 208/230V 60Hz

Description

  • Store whole blood and blood components while meeting the strict AABB, ANRC and FDA requirements with Thermo Scientific Revco Blood Bank Refrigerators, featuring microprocessor control systems, positive airflow systems, industrial-quality cabinet construction with adjustable leveling feet and extra-strength refrigeration compressors designed for industrial, clinical and scientific use.

Features

  • Positive, forced-air circulation maintains temperature uniformity throughout.
  • Directional airflow assures quick recovery after door is opened.
  • High-density, CFC-free urethane foam insulation protects cabinet from ambient temperature fluctuations and minimizes operating costs.
  • Refrigerator includes heavy-duty, industrial-grade, hermetically sealed compressors and automatic condensate removal and defrost systems.

Microprocessor Control

  • Digital electronic technology designed to simplify operation and deliver greater accuracy.
  • Factory-preset temperature setpoint at 4°C.
  • Adjustable temperatures between 1° and 8°C.
  • Icon-based controls.
  • Large, easy-to-read digital display that shows temperature within 0.1°C.
  • Graphic thermometer that shows steady green when temperature is normal, flashes top bar for overtemperature, flashes bulb for undertemperature.
  • Sensor that automatically initiates defrost cycle on demand, eliminating need for a defrost timer, minimizing frost buildup, and optimizing coil efficiency.

Safety and Security

  • Audible/visual warnings for temperature deviations or power failure.
  • Key-operated ON/OFF/ALARM ON master switch for setpoint security.
  • Pushbutton alarm test and remote alarm contacts.
  • Alarm silence, ringback and auto reset.
  • Controller battery backup system.
  • Alarm setpoints preset to warn of temperature deviation at 5.5°C overtemperature, 1.5°C undertemperature.

Construction

  • Cold-rolled steel cabinets with smooth, scratch-resistant painted finish for easy cleaning.
  • Lockable doors self-close against closed-cell peripheral gasket.
  • Dual-pane glass viewing door(s) insulated to prevent condensation.
  • Adjustable, full-extension, stainless-steel drawers.
  • Interior cabinet lights (except undercounter model).
  • Door-activated ON/OFF switch and independent switch on panel.
  • Undercounter model has a solid door and 1in. dia. (2.5cm) access port with plug.

Includes

  • 6" (15.2cm) seven-day temperature chart recorder (except undercounter model—freestanding recorder available separately).

Note

Warranty

  • Valid in the USA only.

Document